CTD Sensor
Wichtige Informationen
Erfordert eine HOBO MX801 oder MX802 Datenloggerdie HOBOconnect-App, und ein kompatibles Mobilgerät oder einen Windows-Computer. Die Systemanforderungen für die App finden Sie am Ende unserer HOBOconnect-Software-Seite.
- 10 für jeweils CHF 1’085.45 kaufen und 3% sparen
- 25 für jeweils CHF 1’063.05 kaufen und 5% sparen
- 50 für jeweils CHF 1’029.50 kaufen und 8% sparen
- 100 für jeweils CHF 984.75 kaufen und 12% sparen
Overview
The CTD sensor is an interchangeable sensor that works with HOBO MX800 Series Water Loggers to measure conductivity, temperature, and water level. With this sensor attached to a HOBO MX800 Series Water Logger, the logger can also calculate specific conductance, salinity, and total dissolved solids (TDS).
The CTD sensor can be attached directly to the fully submersible MX801 model, or attached via cable to the direct read MX802 model, which doesn’t require having to pull the sensor out of the water to download data.
Features
- 4-electrode conductivity sensor provides wide measurement range for monitoring freshwater or saltwater
- Conductivity sensor is centrally located to minimize effects from stilling wells or protective guards
- Durable ceramic water level sensor, located in the sensor tip for accurate measurement of shallow water levels
- Get water level directly with the MX802, no need for post-processing
- 1” diameter for use in narrow wells (with cable to MX802 above the well)
- Temperature sensor placement near the sensor tip for fast response time
- Specially treated stainless steel holds up in saltwater
What's in the box
- CTD Sensor
Specifications
Conductivity Measurements
- Measurement Range
-
Electrical conductivity and specific conductance: 0 to 100,000 µS/cmSalinity using PSS-78: 2 to 42 PSUTotal Dissolved Solids (TDS): 0 to 100,000 mg/L
- Calibrated Range
-
Electrical conductivity: 50 to 80,000 µS/cm5° to 35°C41° to 95°F
- Accuracy
-
Conductivity: +/-2% or 15 µS/cm whichever is greaterSalinity: +/-2% of reading or 0.1 PSU, whichever is greater
- Resolution
-
Conductivity:0.1 µS/cm from 0 to 1,000 µS/cm1 µS/cm from 1,000 to 10,000 µS/cm10 µS/cm from 10,000 to 100,000 µS/cmTDS: 0.1 mg/LSalinity: 0.01 PSU
- Response Time
-
1 second to 90% at a stable temperature
Absolute Pressure and Water Level Measurements W-CTD-01
- Range
-
0 to 207 kPa (0 to 30 psia); approximately 0 to 9 m (0 to 30 ft) of water depth at sea level, or 0 to 12 m (0 to 40 ft) of water at 3,000 m (10,000 ft) of altitude
- Factory Calibrated Range
-
69 to 207 kPa (10 to 30 psia), 0° to 40°C (32° to 104°F)
- Burst Pressure
-
310 kPa (45 psia) or 18 m (60 ft) depth
- Water Level Accuracy*
-
Typical error: ±0.05% FS, 0.5 cm (0.015 ft) waterMaximum error: ±0.1% FS, 1.0 cm (0.03 ft) water
- Raw Pressure Accuracy**
-
±0.3% FS, 0.62 kPa (0.09 psi) maximum error
- Resolution
-
<0.02 kPa (0.003 psi), 0.21 cm (0.007 ft) water
- Pressure Response Time***
-
<1 second to 90% at a stable temperature
Absolute Pressure and Water Level Measurements W-CTD-02
- Range
-
0 to 400 kPa (0 to 58 psia); approximately 0 to 30.6 m (0 to 100 ft) of water depth at sea level, or 0 to 33.6 m (0 to 111 ft) of water at 3,000 m (10,000 ft) of altitude
- Factory Calibrated Range
-
69 to 400 kPa (10 to 58 psia), 0° to 40°C (32° to 104°F)
- Burst Pressure
-
500 kPa (72.5 psia) or 40.8 m (134 ft) depth
- Water Level Accuracy*
-
Typical error: ±0.05% FS, 1.5 cm (0.05 ft) waterMaximum error: ±0.1% FS, 3.0 cm (0.1 ft) water
- Raw Pressure Accuracy**
-
±0.3% FS, 1.20 kPa (0.17 psi) maximum error
- Resolution
-
<0.04 kPa (0.006 psi), 0.41 cm (0.013 ft) water
- Pressure Response Time***
-
<1 second to 90% at a stable temperature
Absolute Pressure and Water Level Measurements W-CTD-03
- Range
-
0 to 850 kPa (0 to 123.3 psia); approximately 0 to 76.5 m (0 to 251 ft) of water depth at sea level, or 0 to 79.5 m (0 to 262 ft) of water at 3,000 m (10,000 ft) of altitude
- Factory Calibrated Range
-
69 to 850 kPa (10 to 123.3 psia), 0° to 40°C (32° to 104°F)
- Burst Pressure
-
1,200 kPa (174 psia) or 112 m (368 ft) depth
- Water Level Accuracy*
-
Typical error: ±0.05% FS, 3.8 cm (0.125 ft) waterMaximum error: ±0.1% FS, 7.6 cm (0.25 ft) water
- Raw Pressure Accuracy**
-
±0.3% FS, 2.55 kPa (0.37 psi) maximum error
- Resolution
-
<0.085 kPa (0.012 psi), 0.87 cm (0.028 ft) water
- Pressure Response Time***
-
<1 second to 90% at a stable temperature
Absolute Pressure and Water Level Measurements W-CTD-04
- Range
-
0 to 145 kPa (0 to 21 psia); approximately 0 to 4 m (0 to 13 ft) of water depth at sea level, or 0 to 7 m (0 to 23 ft) of water at 3,000 m (10,000 ft) of altitude
- Factory Calibrated Range
-
69 to 145 kPa (10 to 21 psia), 0° to 40°C (32° to 104°F)
- Burst Pressure
-
310 kPa (45 psia) or 18 m (60 ft) depth
- Water Level Accuracy*
-
Typical error: ±0.075% FS, 0.3 cm (0.01 ft) waterMaximum error: : ±0.15% FS, 0.6 cm (0.02 ft) water
- Raw Pressure Accuracy**
-
±0.3% FS, 0.43 kPa (0.063 psi) maximum error
- Resolution
-
<0.014 kPa (0.002 psi), 0.14 cm (0.005 ft) water
- Pressure Response Time***
-
<1 second to 90% at a stable temperature
Temperature Measurements (All Models)
- Range
-
-20° to 50°C (-40° to 122°F), non-freezing water
- Accuracy
-
±0.15°C (±0.27°F) from 0° to 50°C (32° to 122°F)
- Resolution
-
0.01°C at 25°C (0.018°F at 77°F)
- Response Time
-
3 minutes to 90% in water (typical)
- Drift
-
<0.1°C (0.18°F) per year
Sensor
- Dimensions
-
2.5 cm (1 inch) diameter, 18.7 cm (7.37 inches) length
- Weight
-
177g6.24 oz
- Wetted Materials
-
Passivated 316 Stainless steel housing rated for use in saltwater, Viton and Buna-N O-rings, PET sensor connectorDepth sensor: ceramicConductivity sensor: PET, platinum plating on electrodesNote: Sensor should be mounted so that it is not in contact with other metals
- Environmental Rating
-
IP68; Waterproof to pressure sensor burst pressure rating
* Water Level Accuracy: With accurate reference water level measurement, known water density, and a stable temperature environment. System Water Level Accuracy equals the sum of the Barometric Water Level Accuracy plus the selected CTD sensor Water Level Accuracy.
** Raw Pressure Accuracy: Absolute pressure sensor accuracy includes all sensor drift, temperature, and hysteresis-induced errors.
*** Changes in Temperature: Allow 20 minutes in water to achieve full temperature compensation of the pressure sensor. There can be up to 0.5% of additional error due to rapid temperature changes.