InTemp Bluetooth Temperature Data Logger with Glycol Bottle (VFC/CDC)
Wichtige Informationen
Für automatische Downloads und Fernwarnungen benötigen die InTemp-Logger das optionale InTemp CX Gateway. MEHR INFORMATIONEN
Overview
InTemp CX402-VFCxxx (with glycol bottle) data loggers monitor temperatures in refrigerators and freezers for clinics participating in the CDC Vaccines for Children program as well as hospitals, clinics, and WHO sites. The loggers communicate wirelessly via Bluetooth to mobile devices. Using the InTemp app, you can easily view data, check logger status, set alarms, and create and share secure PDF reports for streamlined reporting and regulatory compliance. All loggers feature a built-in LCD display to view current and daily min and max temperatures, advanced audible and visual alarm capabilities for notification of temperature excursions, an internal sensor for ambient temperature monitoring, and a 3-year NIST Certification of Calibration (probe only).
With the optional CX Gateway, you can get remote notifications of temperature excursions and automatically download data from loggers within range of the device. Learn more about the CX Gateway here.
Models are available with a 2- or 4-meter external temperature probe attached to a glycol bottle used to insulate the probe and delay thermal response times.
Learn more about our compatible mobile and cloud solutions:
InTempConnect A cloud-based data command-center control for monitoring all loggers throughout your global supply chain
InTemp App A free mobile app that streamlines logger configuration and readout, and automatically uploads reports to InTempConnect
InTempVerify A free mobile app that manages user permissions and the activation of data downloads by anyone—anywhere, anytime
Features
|
|
|
|
|
|
|
|
|
|
Who It's For:
Required to follow Vaccines for Children & CDC requirements, specifically:
- Probe in glycol
- CDC/VFC compliant with 3 year NIST certificate (probe only)
- GDP/GMP compliant
- Frozen to refrigerated monitoring
- For monitoring COVID vaccines
Specifications
External Temperature Probe
- Range
-
-40° to 100°C-40° to 212°F
- Accuracy
-
±1.0°C from -40° to -25°C±1.8°F from -40° to -13°F±0.5°C from -25° to 80°C±0.9°F from -13° to 176°F±1.0°C from 80° to 100°C±1.8°F from 176° to 212°F
- Resolution
-
0.024°C at 25°C0.04°F at 77°F
- NIST Calibration
-
Single point NIST calibration, probe only
- Cable Length
-
2 or 4 meter (6.56 or 13.12 feet) flat ribbon cable
- Probe Dimensions
-
Food-grade stainless steel probe with pointed tip, 53.34 mm (2.1 inches) long, 3.18 mm (0.125 inches) diameter
Temperature Sensor
- Range
-
-30° to 70°C-22° to 158°F
- Accuracy
-
±1.0°C from -30° to -22°C±1.8°F from -22° to -8°F±0.5°C from -22° to 50°C±0.9°F from -8° to 122°F±1.0°C from 50° to 70°C±1.8°F from 122° to 158°F
- Resolution
-
0.024°C at 25°C0.04°F at 77°F
Logger
- Radio Power
-
1 mW0 dBm
- Transmission Range
-
Approximately 30.5 m (100 ft) line-of-sight
- Wireless Data Standard
-
Bluetooth Low Energy (Bluetooth Smart)
- Logger Operating Range
-
-30° to 70°C, 0 to 95% RH (non-condensing)-22° to 158°F, 0 to 95% RH (non-condensing)
- Logging Rate
-
1 second to 18 hours
- Battery Type
-
Two AAA batteries, user replaceable
- Battery Life
-
1 year, typical with logging interval of 1 minute. Faster logging intervals, remaining connected with the InTemp app, excessive report generation, numerous audible alarms, and paging all impact battery life.
- Memory
-
128 KB (84,650 measurements, maximum)
- Full Memory Download Time
-
Approximately 60 seconds; may take longer the farther the device is from the logger
- LCD
-
LCD is visible from 0° to 50°C (32° to 122°F); the LCD may react slowly or go blank in temperatures outside this range
- Dimensions
-
9.4 x 4.5 x 2.59 cm3.7 x 1.77 x 1.02 inches
- Weight
-
90.2 g3.18 oz
- Environmental Rating
-
IP54
Passed RTCA DO160G, part 21H
Documentation
Quick Start for the InTemp CX Temperature Logger (CX40x)
EnglishGerman