![CE Compliance Mark](/images/ce-mark.png)
Data Loggers
Acqua salata
Informazioni importanti
Richiede HOBOware Pro software e un U-DTW-1 Navetta impermeabile o il Base-U-4 Stazione base USB ottica per la configurazione e il trasferimento dei dati. Vedere gli articoli compatibili di seguito.
- Acquista 10 a 1.194,10 CHF ciascuno e risparmia 3%
- Acquista 25 a 1.169,45 CHF ciascuno e risparmia 5%
- Acquista 50 a 1.132,55 CHF ciascuno e risparmia 8%
- Acquista 100 a 1.083,30 CHF ciascuno e risparmia 12%
Overview
A cost-effective data logger for measuring salinity, conductivity, and temperature in saltwater environments with relatively small changes in salinity (±5,000 μS/cm) such as saltwater bays, or to detect salinity events such as upwelling, rainstorm, and discharge events. Data from this logger can be used with data from the HOBO U26 Dissolved Oxygen Logger for monitoring percent dissolved oxygen saturation.
Note: When monitoring salinity in waters with more than ±5,000 uS/cm variation, refer to the accuracy shown in Plot C in the product specifications to determine if the accuracy is acceptable for your needs. Monthly field calibration readings are typically required to achieve the best accuracy.
Features
- Non-contact sensor provides long life
- Easy access to sensor for cleaning and shedding air bubbles
- Two user-selectable ranges provide measurements from 100 to 55,000 uS/cm.
- Delrin housing will not corrode in saltwater
- HOBOware Pro software enables start/end-point calibration to compensate for any fouling and provides easy conversion to specific conductance and salinity using the Practical Salinity Scale 1978 (PSS-78)
- USB optical interface provides high-speed, reliable salinity data offload in wet environments
- Compatible with the HOBO Waterproof Shuttle for easy and reliable data retrieval and transport
- Non-contact sensor with titanium pentoxide coating keeps the sensor from coming in contact with the water, thereby preventing tarnishing or corrosion associated with traditional electrode sensors; sensor coating is inert, so fouling can be wiped off easily
What's in the box
- HOBO U24-002-C Conductivity Data Logger
- Communications window protective cap
Specifications
- Measurements
-
Actual Conductivity, Temperature, Specific Conductance at 25C (calculated), Salinity (calculated using PSS-78, the Practical Salinity Scale 1978)
- Conductivity Measurement Ranges
-
Low range: 100 to 10,000 uS/cmHigh range: 5000 to 55,000 uS/cm
- Conductivity Calibrated Range - Temperature Range
-
5° to 35°C41° to 95°F
- Conductivity Extended Ranges
-
Low range: 100 to 10,000 uS/cmHigh range: 5000 to 55,000 uS/cm(readings below these ranges reported as 0)
- Temperature Measurement Range
-
-2° to 36°C28° to 97°F
- Specific Conductance Accuracy (in Calibrated Range using Conductivity Assistant and Calibration Measurements)
-
Low range: 3% of reading or 50 µS/cm, whichever is greaterHigh range: 5% of reading, in waters within a range of ±3,000 µS/cm; waters with greater variation can have substantially greater error
- Conductivity Resolution
-
2 uS/cm (typical)
- Temperature Accuracy
-
0.1°C (0.2°F) at 25°C (77°F)
- Temperature Resolution
-
0.01°C0.02°F
- Conductivity drift
-
Up to 12% sensor drift per month, exclusive of drift from fouling. Monthly start- and end-point calibration should be used with the HOBOware Conductivity Assistant to achieve the specified Specific Conductance accuracy.
- Response time
-
1 second to 90% of change (in water)
- Operating range
-
-2° to 36°C (28° to 97°F) – non-freezing
- Memory
-
18,500 temperature and conductivity measurements when using one conductivity range; 11,500 sets of measurements when using both conductivity ranges (64 KB total memory)
- Sample rate
-
1 second to 18 hrs, fixed or multiple-rate sampling with up to 8 user-defined sampling intervals
- Clock accuracy
-
+/- 1 minute per month
- Battery life
-
3 years (@ 1 min logging)
- Maximum depth
-
70 m225'
- Weight
-
193 gm (6.82 ounces), buoyancy in freshwater: -59.8 gm (-2.11 ounces)
- Size
-
3.18 cm diameter x 16.5 cm, with 6.3 mm mounting hole (1.25" diameter x 6.5", ¼" hole)
- Wetted housing materials
-
Delrin®, epoxy, stainless steel retaining ring, polypropylene, Buna rubber O-ring, titanium pentoxide (inert coating over sensor) – all materials are suitable for long-term use in saltwater.
- Environmental Rating
-
IP68
![CE Compliance Mark](/images/ce-mark.png)
Documentation
Using the Conductivity Assistant: Calculating Specific Conductance with Sea Water - Instructions
EnglishHOBO U24 Conductivity Logger (U24-002-C) User's Manual
EnglishGerman